Tuberculosis IgG/IgM Rapid Detection Kit
Product Description
Product Detail
Intended Use
Summary And Explanation OF The Test
Tuberculosis is a chronic infectious disease caused by Mycobacterium tuberculosis infection with an incubation period of 4 to 8 weeks, 80% of which occur in the lungs. The disease is mainly transmitted through the respiratory tract. When the patient coughs, sneezes, speaks loudly or spit, the droplets with Mycobacterium tuberculosis are excreted from the body, forming microscopic droplets that float in the air and are inhaled by others to cause infection.
Model:Ag, IgM, IgG, IgM/IgG,IgM and IgG, Ag and IgM/IgG,Ag and IgM and IgG
Materials Provided:
This kit adopts colloidal gold-immunochromatography assay (GICA).
The test card contains:
1. Colloidal gold-labeled antigen and quality control antibody complex.
2. Nitrocellulose membranes immobilized with two test lines (IgG line and IgM line) and one quality control line (C line).
When an appropriate amount of sample is added to the sample well of the test card, the sample will move forward along the test card under capillary action.
If the sample contains an IgG/IgM antibody of Tuberculosis, the antibody will bind to the colloidal gold-labeled Tuberculosis antigen, and the immune complex will be captured by the monoclonal anti-human IgG/IgM antibody immobilized on the nitrocellulose membrane to form a purple/red T line , showing that the sample is positive for IgG/IgM antibody.
Test Procedure
Step1: Allow the test device, buffer, specimen to equilibrate to room temperature (15-30℃) prior to testing.
Step2: Remove the test device from the sealed pouch. Place the test device on a clean, flat surface.
Step3: Label the device with specimen number.
Step4: Using a Disposable Dropper, transfer serum, plasma or whole blood. Hold the dropper vertically and transfer 1 drop of specimen (approximately 10μl) to the specimen well(S) of the test device, and immediately add 2 drops of test buffer (approximately 70-100μl). Make sure there are no air bubbles.
Step5: Set up a timer. Read the results in 15 minutes.
Do not interpret the result after 20 minutes. To avoid confusion, discard the test device after interpreting the result. If you need to store it for a long time, please take a photo of the result.